Effect of porous orthopaedic implant material and structure on load sharing with simulated bone ingrowth: A finite element analysis comparing titanium and PEEK.

Icon for Elsevier Science Related Articles

Effect of porous orthopaedic implant material and structure on load sharing with simulated bone ingrowth: A finite element analysis comparing titanium and PEEK.

J Mech Behav Biomed Mater. 2018 Apr;80:68-76

Authors: Carpenter RD, Klosterhoff BS, Torstrick FB, Foley KT, Burkus JK, Lee CSD, Gall K, Guldberg RE, Safranski DL

Abstract
Osseointegration of load-bearing orthopaedic implants, including interbody fusion devices, is critical to long-term biomechanical functionality. Mechanical loads are a key regulator of bone tissue remodeling and maintenance, and stress-shielding due to metal orthopaedic implants being much stiffer than bone has been implicated in clinical observations of long-term bone loss in tissue adjacent to implants. Porous features that accommodate bone ingrowth have improved implant fixation in the short term, but long-term retrieval studies have sometimes demonstrated limited, superficial ingrowth into the pore layer of metal implants and aseptic loosening remains a problem for a subset of patients. Polyether-ether-ketone (PEEK) is a widely used orthopaedic material with an elastic modulus more similar to bone than metals, and a manufacturing process to form porous PEEK was recently developed to allow bone ingrowth while preserving strength for load-bearing applications. To investigate the biomechanical implications of porous PEEK compared to porous metals, we analyzed finite element (FE) models of the pore structure-bone interface using two clinically available implants with high (> 60%) porosity, one being constructed from PEEK and the other from electron beam 3D-printed titanium (Ti). The objective of this study was to investigate how porous PEEK and porous Ti mechanical properties affect load sharing with bone within the porous architectures over time. Porous PEEK substantially increased the load share transferred to ingrown bone compared to porous Ti under compression (i.e. at 4 weeks: PEEK = 66%; Ti = 13%), tension (PEEK = 71%; Ti = 12%), and shear (PEEK = 68%; Ti = 9%) at all time points of simulated bone ingrowth. Applying PEEK mechanical properties to the Ti implant geometry and vice versa demonstrated that the observed increases in load sharing with PEEK were primarily due to differences in intrinsic elastic modulus and not pore architecture (i.e. 4 weeks, compression: PEEK material/Ti geometry = 53%; Ti material/PEEK geometry = 12%). Additionally, local tissue energy effective strains on bone tissue adjacent to the implant under spinal load magnitudes were over two-fold higher with porous PEEK than porous Ti (i.e. 4 weeks, compression: PEEK = 784 ± 351 microstrain; Ti = 180 ± 300 microstrain; and 12 weeks, compression: PEEK = 298 ± 88 microstrain; Ti = 121 ± 49 microstrain). The higher local strains on bone tissue in the PEEK pore structure were below previously established thresholds for bone damage but in the range necessary for physiological bone maintenance and adaptation. Placing these strain magnitudes in the context of literature on bone adaptation to mechanical loads, this study suggests that porous PEEK structures may provide a more favorable mechanical environment for bone formation and maintenance under spinal load magnitudes than currently available porous 3D-printed Ti, regardless of the level of bone ingrowth.

PMID: 29414477 [PubMed – in process]

]]>

Deformation and fatigue of tough 3D printed elastomer scaffolds processed by fused deposition modeling and continuous liquid interface production.

Icon for Elsevier Science Related Articles

Deformation and fatigue of tough 3D printed elastomer scaffolds processed by fused deposition modeling and continuous liquid interface production.

J Mech Behav Biomed Mater. 2017 11;75:1-13

Authors: Miller AT, Safranski DL, Wood C, Guldberg RE, Gall K

Abstract
Polyurethane (PU) based elastomers continue to gain popularity in a variety of biomedical applications as compliant implant materials. In parallel, advancements in additive manufacturing continue to provide new opportunities for biomedical applications by enabling the creation of more complex architectures for tissue scaffolding and patient specific implants. The purpose of this study was to examine the effects of printed architecture on the monotonic and cyclic mechanical behavior of elastomeric PUs and to compare the structure-property relationship across two different printing approaches. We examined the tensile fatigue of notched specimens, 3D crosshatch scaffolds, and two 3D spherical pore architectures in a physically crosslinked polycarbonate urethane (PCU) printed via fused deposition modeling (FDM) as well as a photo-cured, chemically-crosslinked, elastomeric PU printed via continuous liquid interface production (CLIP). Both elastomers were relatively tolerant of 3D geometrical features as compared to stiffer synthetic implant materials such as PEEK and titanium. PCU and crosslinked PU samples with 3D porous structures demonstrated a reduced tensile failure stress as expected without a significant effect on tensile failure strain. PCU crosshatch samples demonstrated similar performance in strain-based tensile fatigue as solid controls; however, when plotted against stress amplitude and adjusted by porosity, it was clear that the architecture had an impact on performance. Square shaped notches or pores in crosslinked PU appeared to have a modest effect on strain-based tensile fatigue while circular shaped notches and pores had little impact relative to smooth samples. When plotted against stress amplitude, any differences in fatigue performance were small or not statistically significant for crosslinked PU samples. Despite the slight difference in local architecture and tolerances, crosslinked PU solid samples were found to perform on par with PCU solid samples in tensile fatigue, when appropriately adjusted for material hardness. Finally, tests of samples with printed architecture localized to the gage section revealed an effect in which fatigue performance appeared to drastically improve despite the localization of strain.

PMID: 28689135 [PubMed – indexed for MEDLINE]

]]>